Catheter securement systems: comparison of two investigational devices to a sutureless securement device, a securement dressing, and sutures in a pig model

Abstract

Background

Catheter securement is critical for the success of infusion therapy and to prevent complications. Our purpose was to compare the strength of catheter securement achieved with two investigational adhesive securement devices to two securement products and also to sutures using an in vivo animal model.

Methods

Twenty-five live pigs were prepared for aseptic abdominal surgery. Four central venous catheters were inserted per animal into the epigastric veins and secured with four of the five securement systems studied, following a balanced incomplete randomized block design. A peak axial pull force test method was used to measure the force required to dislodge the catheter 1 cm from the insertion site and/or cause failure of the device and/or dressing. This pull test was done 10 min after device application, per constraints of the animal model. Comparison analysis was carried out using a mixed effects model with pig, sample, and sample location as factors. Non-inferiority testing was carried out using 95 % confidence intervals with a margin of 4.52 N or 1 lb (454 g). Tukey’s method was used to adjust for multiple pairwise comparisons.

Results

Results showed that the two investigational devices displayed the highest mean peak axial pull forces (40–41 N) and were significantly better than sutures (28 N, p < 0.0001) and the securement dressing (17 N, p < 0.0001) and non-inferior to the securement device (37 N) in this test. The securement device required a higher mean peak axial pull force than sutures (p = 0.0007) and the securement dressing (p < 0.0001) for failure to occur. Finally, there was also a statistical difference between sutures and the securement dressing, with sutures requiring a higher mean peak axial pull force for catheter dislodgement than the securement dressing (p < 0.0001).

Conclusion

The two investigational devices appear to be a promising alternative for catheter securement, superior to sutures and the securement dressing, and non-inferior to the securement device.

Background

Catheter securement is essential to the success of short-term (<10 days), non-tunneled central venous catheter (CVC) infusion therapy. These short-term CVCs provide an avenue for various therapies, namely the delivery of medications and solutions that cannot be safely infused into the peripheral circulation, fluid resuscitation, repeated blood product replacement, short-term hemodialysis, and the monitoring of hemodynamic status [1]. If these catheters are not effectively secured to the skin, they can become displaced, causing a wide range of complications and safety issues [2, 3]. The various factors affecting catheter securement have been recently reviewed [4].

Catheter securement can be accomplished using several different methods. The most frequently used method of securing short-term CVCs to the skin is silk or nylon sutures [2, 5, 6]. However, the use of sutures is associated with needlestick or sharp injuries to the health-care worker, as well as patient discomfort and the consequences related to disrupting the patient’s skin barrier [7]. To address these issues, manufactured catheter stabilization devices have entered the market in the last decade and are gaining in acceptance to provide securement [8]. Since 2006, various guidelines recommend the use of such securement devices [1, 810]. A recent literature review also found that securement with adhesive systems is more effective than suturing to reduce blood stream infections [11]. Yet, changing practice in health care can take a long time and physicians are likely to continue using sutures for securing CVCs until a larger body of evidence exists for an effective alternative.

To generate evidence comparing sutureless catheter securement methods to sutures, a pig model was used so catheters could be inserted into a dynamic system while avoiding the ethical issues associated with the use of a human model. A peak axial pull force test method was used to determine the relative strength of each securement system in preventing dislodgement under a constant axially directed force. The objective of this study was to explore the catheter securement performance of two novel sutureless systems compared to commercially available sutureless catheter securement systems and sutures. We hypothesized that the novel securement systems would be at least non-inferior to all of the catheter securement systems tested.

Methods

Materials

The following securement systems were applied to the inserted catheters per randomization schedule and manufacturer’s application instructions. The samples were allowed to dwell on the skin for 10 min prior to testing.

Test materials (commercially available at the time of the experiments):

  • Securement device: StatLock® stabilization device (Bard PICC Plus Stabilization Device, Item #VPPCSP, Bard Medical, Covington, GA, USA), covered with a bordered IV cover dressing (3M™ Tegaderm™ Transparent Film Dressing with Border cat.1655, 3M Company, St. Paul, MN, USA).

  • Securement dressing: SorbaView® SHIELD dressing (Cat. # SV353 UDT, Centurion, Williamston, MI, USA).

  • Sutures: 3–0 silk suture (Ethicon, Somerville, NJ, USA), covered with a bordered IV cover dressing (3M™ Tegaderm™ Transparent Film Dressing with Border cat.1655, 3M Company, St. Paul, MN, USA).

Investigational materials

  • Investigational device 1: 3M™ PICC/CVC Securement device + 3M™ Tegaderm™ IV Advanced Securement Dressing (Cat. #1837-2100, 3M Company, St. Paul, MN, USA): a catheter securement device coated with a silicone adhesive and a plastic retainer used to stabilize the catheter lumens, covered with a bordered dressing. This product was under development at the time of this experimentation and is now commercially available.

  • Investigational device 2: 3M™ PICC/CVC Securement device + 3M™ Tegaderm™ CHG Chlorhexidine Gluconate IV Securement Dressing (Cat. # 1877-2100, 3M Company, St. Paul, MN, USA): identical to device 1 but covered with a bordered dressing containing a chlorhexidine gluconate (CHG) gel pad.

Animal model and catheter insertion technique

This study was approved by the contract lab (Synchrony Labs) institutional animal care and use committee and animal care complied with the Guide for the Care and Use of Laboratory Animals and the Animal Welfare Act (9CFR). A pig model was developed by adapting a validated internal test method involving un-inserted catheters placed on the back of humans. This method was modified for use on the pig to include insertion of catheters into the epigastric veins. The investigators were first trained on the validated method prior to proceeding with the animal experiments described below. Training entailed generating peak axial pull force measurements of cut (or un-inserted) catheters placed on the back of a pig using four out of the five study materials (sutures excluded).

Pig model

The study used 26 large female farm pigs (48 ± 3 kg) obtained from Palmetto Research Swine (Reevesville, SC, USA) and was completed in the surgical suite of a facility specializing in pre-clinical models (Synchrony Labs, Durham, NC, USA). The pigs were sedated with Telazol® (4–6 mg/kg) prior to the procedure, and general anesthesia was induced using 5 % isoflurane dispensed through a face mask and maintained with 2–2.5 % isoflurane via endotracheal tube. Animals were placed on the surgery table in dorsal recumbence. Standard ECG leads were applied, and a venous catheter was placed for infusion of fluids in the ear, a location away from the test sites. The animals were intubated, anesthetized, and ventilated prior to testing and were prepared for aseptic abdominal surgery using a standard procedure. The hair on the abdomen was removed using an animal clipper with a #40 blade, and the area was washed with water only. The abdomen was then wiped with 70 % isopropyl alcohol (IPA). Each insertion site on the abdomen was scrubbed with a 10.5 ml, clear ChloraPrep® Patient Preoperative Skin Preparation (CareFusion, San Diego, CA, USA), a 2 % chlorhexidine gluconate/70 % isopropyl alcohol formulation. The ChloraPrep® was allowed to air-dry for at least 5 min to ensure optimal adhesion of the products to skin. In the case of the securement device, the skin was further prepped with the skin protectant included with the device and allowed to dry. During the procedure, animals were monitored for the presence of spontaneous movement and blood oxygen saturation and if needed, appropriate adjustments to anesthesia were made. At the end of the procedure, the animals were given a dose of 40 mEq IV of potassium chloride (KCl) and removed from the ventilator following the American Veterinary Medical Association guidelines on euthanasia [12].

Catheter insertion

A triple lumen 7 Fr., 16-cm-long Arrow®-Howe’s Multi-Lumen Central Venous Catheter (Teleflex® Inc., Research Triangle Park, NC, USA) was aseptically inserted into one of four insertion sites of the superficial epigastric veins and spaced to allow one catheter per location in the right and left upper abdomen and right and left lower abdomen. Insertions were performed using a longitudinal “cut-down” technique to surgically expose the epigastric vein and insert the catheter into the vein under direct vision. The catheter tip was directed toward the animal’s head so that the catheter lumens could be directed caudally.

In order to provide enough space between securement systems, only two catheters were inserted at one time and placed diagonally from one another, e.g., right cranial and left caudal aspects of the abdomen as illustrated in Fig. 1. The skin where each catheter exited was marked with an indelible skin marker. After the first set of two catheters and securement systems were tested, they were pulled out and the second set of two catheters and securement systems were positioned and tested (left cranial, right caudal).

Fig. 1
figure 1

Schematic view of the pig abdomen with two catheters inserted (right cranial, left caudal)

Application of the various catheter securement systems

Figure 2 provides illustrations of the various devices and how they were applied, as modeled on a flat surface for better photographic results. For each device, the first photograph shows the fixation device itself and the second photograph illustrates the device covered with the appropriate dressing. Each device that was not a dressing was covered to comply with its instructions for use to ensure the most clinically relevant test result. For the investigational devices, the first photo shows the fixation device itself, the second photo shows it covered with the dressing without the CHG pad, and the third photo shows the fixation device covered with the CHG dressing.

Fig. 2
figure 2

Application of the investigational materials. a Securement device. b Securement dressing. c Sutures. d Investigational devices 1 and 2. For each device, the first photo shows the fixation device itself, and the second photo shows the final result after the device is covered with the appropriate dressing. For the investigational devices, the first photo shows the fixation device itself, the second photo shows it covered with the dressing without the CHG pad, and the third photo shows the fixation device covered with the CHG dressing

For the securement device (Fig. 2a) and the securement dressing (Fig. 2b), the manufacturer’s instructions were followed. The securement device was covered by a bordered film dressing (3M™ Tegaderm™ Transparent Film Dressing with Border).

Suturing technique (Fig. 2c): To anchor the catheter to the skin, a 3–0 silk suture on a curved needle was first passed through approximately 5-10 mm of skin under each of the two catheter eyelets. After a square knot was tied, the suture was then threaded through each catheter eyelet and three interlocking square knots were made using a needle driver. A dressing was used to cover the insertion site and sutures.

For the investigational systems (Fig. 2d), which consisted of both a device and a dressing, the following application method was used.

First photo: The device was oriented with arrows pointing toward the insertion site. The catheter was placed into the device, and the lumens were woven under the single plastic arm. A large, notched, film-covered soft cloth tape strip is included in the system. The liner of the attached tape strip was removed, and the lumens were secured to the device base by the tape strip. The device was positioned on the skin at the desired location, and the liners were sequentially removed. Pressure was applied to the device base to establish good adhesion to the skin.

Second photo (investigational device 1): After application of the device, 3M™ Tegaderm™ IV Advanced Securement dressing was applied by peeling the liner and placing the dressing so the transparent film covered the insertion site and the border of the dressing covered the single plastic arm on the device. Firm pressure was applied on the dressing including edges to enhance adhesion to the skin. The frame was removed slowly while smoothing down the dressing edges and pressing from the center toward the edges to enhance adhesion. The sterile tape strip was then applied by removing the liner, grasping the non-adhesive tab, and bending slightly with the thumb. The catheter lumens were lifted and the notch end of the tape strip was applied under the lumens and over the dressing edge. Pressure was applied on the tape strip to enhance adhesion. The frame was then removed from the tape strip.

Third photo (investigational device 2): The dressing used to cover the device is 3M™ Tegaderm™ CHG Chlorhexidine Gluconate IV Securement Dressing.

Catheter preparation for testing

To attach the catheter to the IMASS pull tester, the three lumens were clamped and a loop was created by tying the outer two lumens together in a half knot at their distal ends. To prevent the knot from unraveling, a narrow strip of tape was applied to encircle the knot. The IMASS machine was positioned at the foot of the O.R. table, and the “s” hook on the end of the IMASS connection cable was attached to the catheter loop created earlier. This is illustrated in Fig. 3. Using a bubble level attached to the IMASS connection wire, the height of the IMASS table was adjusted so that the IMASS connection wire was axial (level) to the catheter insertion site from which the catheter was pulled.

Fig. 3
figure 3

Lumens tied in a knot and taped; IMASS hook placement on lumens for pull test

Catheter securement testing

The peak axial pull force test was used to measure catheter securement. In this study, catheter securement was defined as the peak force (N) obtained when either of the following failure events occurred (whichever came first):

  • The catheter moves 1 cm or greater out of the insertion site (this amount of motion is clinically relevant as catheter tip migration from the cavo-atrial junction can lead to significant complications in a clinical setting [9])

  • The dressing and/or device lifts off the skin

  • The catheter pulls out from under the dressing

  • The doors of the securement device open

  • The sutures break

The peak axial pull force was measured using an IMASS Model SP-2100 Slip/Peel Tester (IMASS Inc, Accord, MA, USE). The raw peak axial pull force was collected in the units of gram-force, which was converted to the SI unit of Newton using the following formula: 1 g-force = 0.00980665 Newton (N). For comparison, a force of 40 N corresponds to 9 lbs (4.82 kg) of force. The system is described in Fig. 4. Ten minutes after the application of the securement system, the video recording of the entire pull force profile was started, and the pull force (tensile distraction force) was applied at the rate of 76.2 cm/min (30 in./min). The peak axial pull force measurement was obtained when securement system failure occurred. The peak axial pull force (securement of the catheters) was recorded for each securement system applied to each pig. Video recording of the entire pull force profile was obtained and later reviewed. Once peak axial pull force testing was completed with the first two catheters, those were removed, manual pressure was applied to the sites to control bleeding, and the other two insertion sites were accessed. The peak axial pull test method was then repeated for the other two catheters.

Fig. 4
figure 4

Peak axial pull force test equipment

Statistical analysis

A sample size of 25 (no replicates per product on the pig) was planned in this study based on a power calculation. This sample size would achieve at least 80 % power to detect a difference of 454 gram force or 4.45 N (using an adjusted Tukey test) if the within-pig standard deviation (SD) was at most 420 gram force (4.12 N), assuming 94 % efficiency with the incomplete block design of having five samples to choose from with a maximum of four samples per pig.

Peak axial pull force analysis was conducted using a mixed effects model at significance level of 0.05 (two-sided). Sample and sample location (upper versus lower sites) were used as the fixed effects in the model, while subject (pig identifier) was included in the model as the random effect. The sample-by-location interaction term was tested for significance. Heterogeneity was modeled in which each sample was allowed a different covariance parameter.

Non-inferiority/superiority testing was carried out using the two-sided 95 % adjusted confidence limits and the margin of 4.45 N or 1.0 lb force or 454 gram force. Tukey’s method was used to adjust for multiple pairwise comparisons, which were based on the differences in least squares means.

Results

Reproducibility of model

Protocol deviations with respect to sample placement occurred with two early pigs such that there were only three products placed instead of the expected eight products. An additional pig was utilized to obtain four out of the five missing products. This created an imbalance in the placement of sutures and the investigational device 1 samples. Table 1 displays the number of replicates and location for each system tested. The 26 pigs provided a total of 99 sites. The overall sample size consisted of 20 data points for all samples except sutures in which 19 data points were recorded. Table 2 provides the summary of the data collected for each device tested along with information on the reproducibility of this test method.

Table 1 Number of replicates for each securement device and location on pig
Table 2 Summary statistics of peak axial pull force dataset

The sample-by-location interaction term was non-significant. The final model chosen included just the main effects of sample and location (upper vs lower insertion sites). Based on the final model, the within-pig SD was estimated at 0.99 N and the between-pig SD was estimated between 5.16 and 6.91 N, with the securement dressing having the lowest SD and investigational device 1 having the highest SD.

Performance of the securement systems

All systems had the same mode of failure, which was the movement of the catheter 1 cm or greater out of the insertion site (catheter dislodgement). In one case, a suture broke on one side of the catheter. Figure 5 illustrates graphically the peak axial pull forces for all methods of securement compared in this study.

Fig. 5
figure 5

Mean peak axial pull force for each type of securement device (least squares, mean ± standard error). Samples labeled with the same symbols are not significantly different from each other. Least squares means differ slightly from the unadjusted means displayed in Table 2 due to the overall imbalance of securement devices placed on pigs

Results showed that the two investigational devices displayed the highest mean peak axial pull forces (40–41 N) and were significantly better than sutures (28 N, p < 0.0001) and the securement dressing (17 N, p < 0.0001) and non-inferior to the securement device (37 N) in this test (Fig. 5). Both investigational systems 1 and 2 were estimated at 3.31 N (95 % CI −1.14, +7.76 N) and 4.03 N (95 % CI −0.28, +8.34 N) higher than the securement device, respectively, and proved to be non-inferior to it based on the lower 95 % confidence limit falling within the non-inferior margin (≥4.452 N).

The securement device required a higher mean peak axial pull force than sutures (p = 0.0007) and the securement dressing (p < 0.0001) for failure to occur. Finally, the difference in mean peak axial pull force between sutures and the securement dressing also reached statistical significance (28 versus 17 N, p < 0.0001).

There was no statistical difference between the two investigational devices, which on average differed from each other by less than 1 N (95 % confidence limits −5.05 N, +3.62 N). Both investigational devices had means that were 12.7–13.4 N higher than sutures and 22.9–23.6 N higher than the securement dressing.

Overall, both investigational devices were significantly better than sutures (50 % stronger), significantly better than the securement dressing (140 % stronger), and non-inferior to the securement device in this test method. None of these devices caused damage to the skin of the animals. However, since the animals were under anesthesia during the whole procedure, the catheter pull test was performed shortly after the application of the securement devices and not several days later to simulate clinical wear time, which is a limitation of this model.

Discussion

Our results showed that the two investigational devices displayed the highest mean peak axial pull forces (40–41 N) and were significantly better than sutures (28 N, p < 0.0001) and the securement dressing (17 N, p < 0.0001) and non-inferior to the securement device (37 N) in this test.

A growing body of clinical literature has described the use of securement (or stabilization) devices and dressings for intravenous catheters in order to reduce complications. From this literature, StatLock® appears to be the most commonly studied sutureless catheter securement device for CVCs [1321]; other devices such as the Nexiva™ Closed IV Catheter System with the customized 3M™ Tegaderm™ IV Advanced Securement Dressing, the Sorbaview® dressings, the HubGuard® catheter securement dressings, and the SecurAcath® device have also been the topic of research publications [2226]. Catheter securement reduces phlebitis, infection, catheter migration, and dislodgement (reviewed in [13]). Other authors have published papers stating that securement devices help to reduce catheter-related bloodstream infections (CR-BSIs) and suture-related needlesticks and improve cost-effectiveness in patients with central venous access devices, including peripherally inserted central catheters (PICCs) [7, 11]. The investigational devices we describe here have not been tested for CR-BSI reduction at this time. The effect of securement on catheter complications has been reviewed from a clinical standpoint in [27]. These important clinical studies look at factors such as complication rates, ease of use, dwell time, patient comfort, and cost but do not address the mechanical properties such as pull force to compare the various products since it would be unethical to run such experiments on patients.

Although catheter securement systems have been observed to be very useful in clinical use, the rigorous, quantitative testing of their mechanical performance is still in its infancy. Since it would be unethical to subject patients to the risks of tissue injury and catheter dislodgement, models have been created for this type of testing, and they each have their own limitations. For example, forces required to pull out catheters secured with various tapes and taping methods have been measured with a force transducer using a simulated forearm model consisting in PVC pipe [28]. The same group also conducted experiments on the forearm of human volunteers, with catheters taped to the skin but not inserted into veins. We have also used a similar method to test peak axial pull force on human volunteers, in addition to performing a drop test to simulate a fast pull [29]. Other researchers have used pig skin in vitro to suture a catheter adjacent to the catheter entry site and suspend a 1 lb weight to check the tethering, but they have not specified the duration of weight application and observations [30]. In a different study, postmortem newborn pig hind legs were used to measure the pull force of catheters secured in different ways (standard or bordered polyurethane films, StatLock®, or tissue adhesives) [31]. It was not specified however whether the catheters were inserted in a vein or just in the tissue. They obtained a mean pull force of about 22 N for StatLock® in their model, versus a mean value of about 37 N using our model, indicating differences between the models. They did not use sutures as a comparator. A porcine hind limb model was also used to compare catheterization techniques for peripheral nerve block and the resulting force required to withdraw catheters, but no securement systems were tested in that study [32]. A different group used the proximal end of discarded extremities after below-the-knee amputations to serve as a surrogate tissue for catheter fixation [33]. In this study, a rod was used to tunnel the catheter under the skin to simulate placement into a central vessel. They compared sutures to two types of staples and measured pull forces. The mean peak force required to dislodge sutures was around 41 N (we were around 27 N in our model). Their study did not test adhesive securement systems.

We have used our animal model to compare various adhesive systems to sutures in the same set of experiments. We chose the pig because of the feasibility of inserting catheters in blood vessels of a size comparable to those of humans and because pig skin has been considered the most similar to humans among the non-primate animals for decades in in vivo wound healing research [34, 35]. In addition, we believe that the reaction of pig skin to adhesives is likely to model human skin because both species have a similar total turnover time for the epidermis (30 days for the pig and 26–27 days for humans) [36]. We believe that our model presents an advantage over the other models from the literature, because the catheters were truly inserted into veins of living (anesthetized) animals. The peak axial pull force values measured are therefore likely to better approximate the clinical situation and provide a good basis for the comparison of various securement devices. Since we also generated internal data using the same pull force testing equipment on human volunteers (without insertion of the catheters into veins and without the suture comparison), we are able to comment on the two models. When comparing the pull force data generated in this pig model to the pull force data generated in the human in vivo model, the values were higher in the pig model than in the human model (the difference ranged on average from 4.6 to 9.2 N). Factors related to a catheter being inserted in the veins and the skin of the pig (no moisture or sweating compared to humans) may be contributing to the observed differences in readings. As found in the pig study, the findings in our human study had indicated that the securement dressing required significantly less force to dislodge the catheter than all three of the sutureless securement systems tested, and the three sutureless securement devices showed pull forces that were not different from each other at every time point tested (5 min, 4, and 7 days; manuscript in preparation). Therefore, the comparative performance of the adhesive securement systems tested was the same in both models, confirming the relevance of the pig model in effectively predicting the differences and similarities when measuring peak axial pull force.

This study has some limitations. Since we used a model with animals under anesthesia, the catheter pull test was performed shortly after the application of the securement devices and not several days later to simulate clinical wear time. We also did not torque the catheters or wet the dressings or compare different time points after application. We chose to insert the catheters in epigastric vessels to reduce the number of animals needed compared to using the jugular veins only. We do believe however that our method is informative to compare various devices for their securement performance after application. We chose the Bard StatLock® PICC Plus Stabilization Device as a comparator because we had also used it in our human model without vein insertion and already had comparative data. Moreover, this application is not contraindicated in the instructions for use of this product. We believe our results are relevant for both PICCs and CVCs since our model focuses on the performance of the device when resisting a pull force. For our suture comparator, we selected one type of suture and one fixation method, which may limit the generalizability of the comparison. The possible disadvantages of the new systems are comparable to those of other adhesive securement systems, such as pain or catheter migration while removing or replacing the dressing, and possible skin irritation due to the adhesive on patients with sensitive skin.

Conclusions

In conclusion, the investigational systems were significantly better than sutures and the securement dressing and non-inferior to the securement device in this pig model using a pull force test method performed 10 min after application of the devices. Securement devices specifically designed to house the catheter performed better than a dressing alone, and such devices also performed better than sutures. Clinically, they offer the advantage of avoiding additional skin perforations for patients as well as the risk of needlestick injuries for health-care workers. The investigational sutureless catheter securement systems described here will provide clinicians with viable securement alternatives to sutures and other sutureless catheter securement systems, while offering the advantage of a convenient all-in-one system. Additional clinical benefits have been identified in a clinical user survey study (manuscript in preparation).

The development of test methods that provide evidence not readily obtained in the clinical setting is important to understand the fundamental mechanical properties of these systems and to quantitatively compare their performance. Clinical studies designed to compare the adverse events observed with various sutureless catheter securement systems will be needed to identify the clinically relevant differences between these securement systems.

Abbreviations

CFR:

Code of Federal Regulations

CHG:

chlorhexidine gluconate

CVC:

central venous catheter

ECG:

electrocardiogram

IPA:

isopropyl alcohol

N:

Newtons

PICC:

peripherally inserted central catheter

PVC:

polyvinyl chloride

SD:

standard deviation

References

  1. Bishop L, Dougherty L, Bodenham A, Mansi J, Crowe P, Kibbler C, Shannon M, Treleaven J (2007) Guidelines on the insertion and management of central venous access devices in adults. Int J Lab Hematol 29:261–278

    Article  CAS  PubMed  Google Scholar 

  2. Biggs TC, Mohammed A, Mulvihill D (2012) Importance of securing central venous catheters. Anaesth Intensive Care 40:366–367

    CAS  PubMed  Google Scholar 

  3. Nayeemuddin M, Pherwani AD, Asquith JR (2013) Imaging and management of complications of central venous catheters. Clin Radiol 68:529–544

    Article  CAS  PubMed  Google Scholar 

  4. Macklin D, Blackburn PL (2015) Central venous catheter securement: using the healthcare and technology synergy model to take a closer look. JAVA 20:45–50

    Article  Google Scholar 

  5. Rupp SM, Apfelbaum JL, Blitt C, Caplan RA, Connis RT, Domino KB, Fleisher LA, Grant S, Mark JB, Morray JP, Nickinovich DG, Tung A (2012) Practice guidelines for central venous access: a report by the American Society of Anesthesiologists Task Force on central venous access. Anesthesiology 116:539–573

    Article  PubMed  Google Scholar 

  6. Vinjrayer A, Jefferson P (2004) Securing central venous catheters: a comparison of sutures with staples. Emerg Med J 21:582–583

    Article  Google Scholar 

  7. Rosenthal K (2005) Get a hold on costs and safety with securement devices. Nurs Manage 36:52–54

    Article  PubMed  Google Scholar 

  8. Schears G (2006) New standards to improve catheter stabilization and patient and worker safety. Journal of healthcare risk management: the journal of the American Society for Healthcare Risk Management 26:15–18

    Article  Google Scholar 

  9. Infusion Nurses Society (2011) Infusion nursing standards of practice. J Infusion Nursing 34:1–110

    Article  Google Scholar 

  10. O'Grady NP, Alexander M, Burns LA, Dellinger EP, Garland J, Heard SO, Lipsett PA, Masur H, Mermel LA, Pearson ML, Raad II, Randolph A, Rupp ME, Saint S, Committee HICPAC (Healthcare Infection Control Practices Advisory Committee) (2011) Guidelines for the prevention of intravascular catheter-related infections. Cent Dis Control, p 83. http://www.cdc.gov/hicpac/BSI/BSI-guidelines-2011.html

  11. Petree C, Wright DL, Sanders V, Killion JB (2012) Reducing blood stream infections during catheter insertion. Radiol Technol 83:532–540

    PubMed  Google Scholar 

  12. American Veterinary Medical Association (2013) AVMA Guidelines for the euthanasia of animals: 2013 edition. https://www.avma.org/KB/Policies/Documents/euthanasia.pdf. Accessed on 8 July 2015

  13. Alekseyev S, Byrne M, Carpenter A, Franker C, Kidd C, Hulton L (2012) Prolonging the life of a patient’s IV: an integrative review of intravenous securement devices. Medsurg Nurs 21:285–292

    PubMed  Google Scholar 

  14. Stephenson C (2005) The advantages of a precision-engineered securement device for fixation of arterial pressure-monitoring catheters. JAVA 10:130–132

    Article  Google Scholar 

  15. Frey AM, Schears GJ (2006) Why are we stuck on tape and suture? J Infusion Nursing 29:34–38

    Article  PubMed  Google Scholar 

  16. Schears GJ, Liebig C, Frey AM, Chiavacci R, Harris K, Lin RJ, Costarino AT, Helfaer MA (2000) StatLock catheter securement device significantly reduces central venous catheter complications. Patient Safety Initiative 2000:28–33

    Google Scholar 

  17. Schears GJ (2006) Summary of product trials for 10, 164 patients: comparing an intravenous stabilizing device to tape. J Infus Nurs 29:225–231

    Article  PubMed  Google Scholar 

  18. Bausone-Gazda D, Lefaiver CA, Walters SA (2010) A randomized controlled trial to compare the complications of 2 peripheral intravenous catheter-stabilization systems. J Infus Nurs 33:371–384

    Article  PubMed  Google Scholar 

  19. Royer T (2003) Improving short peripheral IV outcomes: a clinical trial of two securement methods. JAVA 8:45–49

    Article  Google Scholar 

  20. Smith B (2006) Peripheral intravenous catheter dwell times: a comparison of 3 securement methods for implementation of a 96-h scheduled change protocol. J Infus Nurs 29:14–17

    Article  PubMed  Google Scholar 

  21. Yamamoto AJ, Solomon JA, Soulen MC, Tang J, Parkinson K, Lin R, Schears GJ (2002) Sutureless securement device reduces complications of peripherally inserted central venous catheters. J Vasc Interv Radiol 13:77–81

    Article  PubMed  Google Scholar 

  22. Delp J, Hadaway L (2011) New product decisions—the process and outcome for a community health system. JAVA 16:74–84

    Article  Google Scholar 

  23. McNeill EE, Hines NL, Phariss R (2009) A clinical trial of a new all-in-one peripheral-short catheter. JAVA 14:46–51

    Article  Google Scholar 

  24. Flippo PL, Lee J (2011) Clinical evaluation of the Sorbaview SHIELD securement device used on peripheral intravenous catheters in the acute care setting. JAVA 16:95–102

    Article  Google Scholar 

  25. Cordovani D, Cooper RM (2013) A prospective trial on a new sutureless securement device for central venous catheters. Can J Anaesth 60:504–505

    Article  PubMed  Google Scholar 

  26. Egan GM, Siskin GP, Weinmann R, Galloway MM (2013) A prospective postmarket study to evaluate the safety and efficacy of a new peripherally inserted central catheter stabilization system. J Infus Nurs 36:181–188

    Article  PubMed  Google Scholar 

  27. Gabriel J (2010) Vascular access devices: securement and dressings. Nurs Stand 24:41–46

    Article  PubMed  Google Scholar 

  28. Patel N, Smith CE, Pinchak AC, Hancock DE (1995) Evaluation of different methods of securing intravenous catheters: measurement of forces during simulated accidental pullout. Can J Anaesth 42:504–510

    Article  CAS  PubMed  Google Scholar 

  29. 3M PICC CVC Securement Drop Test short. https://www.youtube.com/watch?v=bcYq1T2p5vM. Accessed on 8 July 2015.

  30. Poulton B, Young P (2000) A novel method for epidural catheter fixation. Anaesthesia 55:1141–1142

    Article  CAS  PubMed  Google Scholar 

  31. Simonova G, Rickard CM, Dunster KR, Smyth DJ, McMillan D, Fraser JF (2012) Cyanoacrylate tissue adhesives—effective securement technique for intravascular catheters: in vitro testing of safety and feasibility. Anaesth Intensive Care 40:460–466

    CAS  PubMed  Google Scholar 

  32. Tsui BC, Tsui J (2012) Less leakage and dislodgement with a catheter-over-needle versus a catheter-through-needle approach for peripheral nerve block: an ex vivo study. Can J Anaesth 59:655–661

    Article  PubMed  Google Scholar 

  33. Motonaga GK, Lee KK, Kirsch JR (2004) The efficacy of the arrow staple device for securing central venous catheters to human skin. Anesth Analg 99:1436–1439, table of contents

    Article  PubMed  Google Scholar 

  34. Meyer W, Schwarz R, Neurand K (1978) The skin of domestic mammals as a model for human skin, with special reference to the domestic pig. CurrProblDermatol 7:39–52

    CAS  Google Scholar 

  35. Sullivan TP, Eaglstein WH, Davis SC, Mertz PM (2001) The pig as a model for human wound healing. Wound Repair Regen 9:66–76

    Article  CAS  PubMed  Google Scholar 

  36. Weinstein GD (1966) Comparison of turnover time and of keratinous protein fractions in swine and human epidermis. In: Bustad LK, McClellan RO (eds) Swine in biomedical research. Frayn Printing Co., Seattle, pp 287–297

    Google Scholar 

Download references

Acknowledgements

The authors wish to acknowledge the professional work of the staff at Synchrony Lab. This study was funded by 3M.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Stéphanie F. Bernatchez.

Additional information

Competing interests

All authors are employees of the 3M Company. The experiments were conducted in an independent testing facility (Synchrony Lab).

Authors’ contributions

LFR designed the study protocol, coordinated the interaction with the contract laboratory, and participated in the draft of the introduction section. DPD co-developed the investigational devices and contributed to the design of the study protocol. SAHW conducted the statistical analysis of the data, wrote the statistical methods section, and contributed to writing the results section. SFB conducted the literature review, contributed to the interpretation of the data, and drafted the overall manuscript with responsibility for the “Discussion” section. All authors discussed the interpretation of the data, as well as read and approved the final manuscript.

Authors’ information

All authors are employees of the 3M Company. This study was designed with the goal of establishing an in vivo pre-clinical model to perform quantitative testing of the mechanical performance of catheter securement systems in order to compare devices under development to devices currently used in the clinical setting. LFR is a certified registered nurse infusionist; DPD is a chemical engineer and a Six Sigma master black belt; SAHW is a biostatistician; and SFB has a PhD in Interdisciplinary Sciences.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Rutledge, L.F., DeCabooter, D.P., Walters, SA.H. et al. Catheter securement systems: comparison of two investigational devices to a sutureless securement device, a securement dressing, and sutures in a pig model. ICMx 3, 24 (2015). https://doi.org/10.1186/s40635-015-0060-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40635-015-0060-3

Keywords