Reliability of the nitrogen washin-washout technique to assess end-expiratory lung volume at variable PEEP and tidal volumes

Abstract

Background

End-expiratory lung volume measurement by the nitrogen washin-washout technique (EELVWI-WO) may help titrating positive end-expiratory pressure (PEEP) during acute respiratory distress syndrome (ARDS). Validation of this technique has been previously performed using computed tomography (EELVCT), but at mild PEEP levels, and relatively low fraction of inspired oxygen (FiO2), which may have insufficiently challenged the validity of this technique. The aims of this study were (1) to evaluate the reliability of EELVWI-WO measurements at different PEEP and VT during experimental ARDS and (2) to evaluate trending ability of EELVWI-WO to detect EELV changes over time.

Methods

ARDS was induced in 14 piglets by saline lavage. Optimal PEEP was selected during a decremental PEEP trial, based on best compliance, best EELVWI-WO, or a PEEP-FiO2 table. Eight VT (4 to 20 mL · kg-1) were finally applied at optimal PEEP. EELVWI-WO and EELVCT were determined after ARDS onset, at variable PEEP and VT.

Results

EELVWI-WO underestimated EELVCT with a non-constant linear bias, as it decreased with increasing EELV. Limits of agreement for bias were ±398 mL. Bias between methods was greater at high PEEP, and further increased when high PEEP was combined with low VT. Concordance rate of EELV changes between consecutive measurements was fair (79%). Diagnostic accuracy was good for detection of absolute EELV changes above 200 mL (AUC = 0.79).

Conclusions

The reliability of the WI-WO technique is critically dependent on ventilatory settings, but sufficient to accurately detect EELV change greater than 200 mL.

Background

Acute respiratory distress syndrome (ARDS) is characterized by a major decrease in lung aerated volume. End-expiratory lung volume measurement by the nitrogen washin-washout technique (EELVWI-WO) [1] is available at the bedside from an ICU ventilator, and may help titrating PEEP during mechanical ventilation of ARDS patients. Validation of this technique has been previously performed in mechanically ventilated patients using computed tomography (CT) as gold standard [2], but at relatively low PEEP levels (5 cm H2O), low fraction of inspired oxygen (FiO2) and respiratory rate (RR), and with tidal volume (VT) 8 ± 1 mL · kg-1 in the upper range of current experts’ recommendations for ARDS management [3]. Such ventilatory settings may have insufficiently challenged the validity of this technique, which requires a constant inhomogeneity in alveolar gas throughout the measurement, and may be less precise at FiO2 greater than 0.7 [1].

Furthermore, the WI-WO technique is particularly suitable for repeated EELV assessment, and hence to identify EELV trends, but has never been formally validated for this purpose.

The aims of this study were to evaluate (1) the reliability of EELVWI-WO measurement at variable PEEP and VT, at high RR and FiO2 during experimental ARDS, using CT as a reference and (2) the trending ability of WI-WO technique to detect change in EELV associated with PEEP and VT variations.

Methods

This study was approved by our Institutional Review Board for the care of animal subjects (Comité d’experimentation animale de l′université Lyon I), and carried out in 14 pigs (28 ± 2 kg).

Animal preparation

Pigs were anesthetized with propofol and fentanyl, tracheotomized and mechanically ventilated in volume-controlled mode, with constant inspiratory flow, VT 10 mL · kg-1, inspired fraction of oxygen (FiO2) 0.21, zero end-expiratory pressure, and RR adjusted to achieve normocapnia using Engström Carestation® ventilator (General Electric Healthcare, Madison, WI, USA). Muscle relaxation was obtained with pancuronium bromide. Right jugular vein was cannulated with a 3-lumen 8.5-Fr catheter for drug administration. Carotid artery was cannulated with an 8.5 Fr catheter. FiO2 was increased to 1 at the end of animal preparation.

Measurements

Air flow was measured using a small volume pneumotachograph (PN 281637, Hamilton medical AG, Bonaduz, Switzerland). Pressure at the airway opening was measured using a connecting tube with lateral aperture connected between the endotracheal tube and the pneumotachograph. Signals of arterial blood pressure, pressure at the airway opening, and air flow were read by transducers (Becton Dickinson, Sandy, UT, USA), connected to an A/D card (MP 100; Biopac Systems, Santa Barbara, CA, USA), acquired at 200 Hz and analyzed with Acknowledge® software (Biopac Systems, Santa Barbara, CA, USA). Tracheal pressure was measured through an air filled catheter introduced down the endotracheal tube, positioned 2 cm distal to the tube tip, and connected to the ventilator, to obtain alveolar pressure [4].

EELVWI-WO was assessed by the ventilator, by using the nitrogen washout/washin technique [1], from continuous measurement of end-tidal O2 and CO2 during a 0.1 change of FiO2 using pediatric sensors (Pedi-lite+, Dahtex-Ohmeda Inc, Madison, WI, USA). The average value of the washout and washin measurements during 1 to 0.9 and 0.9 to 1 FiO2 changes was given by the ventilator.

EELVCT was calculated using lung CT, as previously described [2]. CT calibration using the manufacturer phantom was performed before each CT study. The CT scanner (Biograph mCT/S, Siemens, Munich, Germany) was set as follows: interval 1 mm, voltage 120 kV, pitch 1.2 mm, and field of view 300 mm. Whole lung CT images were taken during 15 s end-expiratory. CT raw data were reconstructed as 1-mm-thick contiguous slices using a medium smooth filter (B31f). Image segmentation was manually performed over the whole lung using Turtleseg® software [5, 6] (http://www.turtleseg.org). Gas volume in each lung voxel was computed from the CT number using the following formulas [2]:

Gas volume = CTnumber - 1 , 000 × voxel volume for lung voxels with CT number 0 .

Gas volume = 0 for lung voxels with CT number > 0.

EELVCT was computed as the sum of gas volume in all the voxels defined by lung segmentation.

Expected EELV on zero end-expiratory pressure was deemed as 33 mL · kg-1 body weight as previously published in normal anesthetized pigs [7].

Protocol

ARDS was performed by saline lavage at ventilatory settings mentioned above. Intra-tracheal instillations of 1,000 mL aliquots of 0.9% sodium chloride warmed at 37°C were repeated until PaO2/FiO2 ratio was <100 mmHg. RR may be increased up to 35 breaths per min to maintain pH above 7.20, then kept constant except at the end of experiment, where at the highest VT, it was decreased to maintain peak airway pressure below 100 cm H2O.

Then, PEEP was set to 20 cm H2O, VT to 6 mL · kg-1, and a recruitment maneuver was performed by applying a continuous airway pressure of 40 cm of H2O over 40 s. A decremental PEEP trial was then performed from 20 to 2 cm H2O by 2 cm H2O steps of 10 min each. At the end of the decremental PEEP trial, animals were randomized into three PEEP groups, for which PEEP level was set according to either highest compliance (n = 4), or highest EELVWI-WO (n = 5), or PEEP-FiO2 table (n = 4) [8]. This randomization was used to deliver a wide PEEP range during the final part of the study, in order to obtain multiple combinations of PEEP and VT, so as to perform a multivariate analysis adjusted for PEEP and VT. One pig died just after the PEEP trial before randomization, and was kept in the final analysis. The selected PEEP was applied for 1 h, and VT was adjusted to maintain plateau pressure of the respiratory system ≤30 cm of H2O.

After 1 h of applied selected PEEP, eight levels of VT (4, 5, 6, 7, 8, 10, 15, 20 mL · kg-1) ranging from 100 to 625 mL, were applied for 2 min leaving PEEP level unchanged. EELVWI-WO and EELVCT were measured immediately after ARDS onset, at the end of each PEEP step during the PEEP trial, and at the end of each VT change. A 15-s end-inspiratory pause was performed to check the absence of air leak in each experimental condition. Some experimental conditions were not available since pneumothorax occurred in several pigs at high VT or since EELVWI-WO values were lacking for technical reasons, ending up in 218 data points in final analysis (see Additional file 1: Table S1 for description of lacking data points).

Statistical analysis

Statistical analyses were performed using R software [9], with packages nlme [10], MethComp [11], pROC [12], OptimalCutpoints [13], and multcomp [14]. Values were expressed as mean ± standard deviation (SD). The level of statistical significance was set below 0.05.

EELVWI-WO and EELVCT were compared using a linear mixed-effects model, and Bland and Altman representation [15]. Limits of agreement were computed using alternating regression [16] since bias was non-constant and the experimental design involved repeated measurements.

To control for an effect of confounding variables on the bias, a linear mixed-effects model was built using PEEP, VT, EELVCT at ARDS onset and their interactions as factors with fixed effect, pigs as factor with random effect [17], and bias as dependent variable. Model simplification was performed using a backward stepwise algorithm.

Percentage error was computed as × SDBias/meanEELV[18]. As percentage error was not reported in the two previously published studies that compared EELVWI-WO and EELVCT[2, 19], Cartesian data of these studies were reanalyzed, being uplifted using a scientific program allowing extraction of individual data points from a digitalized graph (DataThief III®[20]) as follows: a digital copy of each regression plot was analyzed with DataThief from the portable document format file of the journal articles, and the extracted data were exported as two columns of X-Y coordinates, with each row representing an extracted data point, allowing computation of percentage error of each study.

Changes in EELV between consecutive measurements were computed for EELVWI-WO (ΔEELVWI-WO) and EELVCT (ΔEELVCT). Ability of the WI-WO technique to track changes in EELV was assessed using four-quadrant and polar plots. The four-quadrant plot relates ΔEELVWI-WO and ΔEELVCT, with upper right and lower left quadrants being quadrants of agreement (in which both EELVWI-WO and EELVCT have the same directional changes) and lower right and upper left quadrants being quadrants of disagreement (in which EELVWI-WO and EELVCT have opposite directional changes). Concordance rate was defined as the percentage of data points falling into one of the two quadrants of agreement, expressed as a percentage of the total number of data points [21]. The main drawback of the four-quadrant plot is the lack of quantification of the distance between each data point and the line of identity, leading to the development of polar plot analysis [21]. Polar plot is obtained by a 45° clockwise rotation of the four-quadrant plot, changing the dimensions of the radius to mean ΔEELV [22], lining up the line of identity along the horizontal axis. Data points with positive and negative directional changes are located on the right and the left side of the polar plot, respectively, and the polar angle represents the angle of each data point with line of identity. A 0° polar angle depicts a perfect agreement between ΔEELVWI-WO and ΔEELVCT, while polar angles in the range 45° to 135° and 225° to 315° depict disagreement between directional changes of EELVWI-WO and EELVCT. The following variables are computed from polar plots: (1) angular bias as the mean angle between all data points and polar axis [21], reflects the difference in calibration between the reference and test methods; (2) radial limits of agreement as the radial sector containing 95% of the data points, after conversion of negative deflections to positive ones, is a polar version of the 95% confidence limits and is similar to the limits of agreement in Bland and Altman analysis [21].

Bias and angular bias were compared to zero using Mann-Whitney U test. Multiple comparisons were performed with Dunnett’s test using PEEP 0 as a reference.

The ability of the WI-WO technique to detect a change in EELV greater than 100, 150, 200, 250, and 300 mL was tested by computations of area under receiver operating characteristic (AUC) curve. The optimal cut-off points were computed using the Youden J statistic.

Results

Ventilatory settings and arterial blood gases during the whole experiment are reported in Additional file 2: Table S2. Figure 1 depicts the evolution of EELVWI-WO and EELVCT over time. Immediately after ARDS onset, EELVCT was very low (236 ± 143 mL (25% ± 15% of its theoretical value, range 104 to 668 mL)), and increased to 1,206 ± 185 mL (range 957 to 1,528 mL) at PEEP 20.

Figure 1
figure 1

EELV WI-WO and EELV CT at each experimental condition (upper panel), with corresponding RR, PEEP, and V T (lower panel). Values are mean ± standard deviation. ARDS, experimental acute respiratory distress syndrome; EELVWI-WO, end-expiratory lung volume assessed with the nitrogen washin-washout technique; EELVCT, end-expiratory lung volume assessed by computed tomography; optimal PEEP, optimal PEEP level according to one of the three methods (see text for details); PEEP, positive end-expiratory pressure; RR, respiratory rate; VT, tidal volume.

EELCCT and EELVWI-WO values were very close at PEEP 0, but their difference progressively increased with the PEEP level.

Comparison of EELVWI-WO and EELVCT

EELVWI-WO and EELVCT were significantly correlated (R2 = 0.63, p < 0.001). The regression equation between EELVWI-WO and EELVCT, had an intercept of 96 mL (p < 0.001) and a slope of 0.58 mL-1 (p < 0.001; Figure 2). Bland and Altman representation exhibited a non-constant bias, decreasing toward more negative values as mean EELV increased (Figure 3). The difference between EELVWI-WO and EELVCT was related to their mean value by the following equation:

Figure 2
figure 2

Relationship between EELV WI-WO and EELV CT . Each symbol represents a concomitant measurement of end-expiratory lung volume assessed with either nitrogen washin-washout technique (EELVWI-WO) or computed tomography (EELVCT). Solid line is the regression line. Dashed line is the line of identity.

Figure 3
figure 3

Bias and limits of agreement between EELV CT and EELV WI-WO , using Bland and Altman representation. Each symbol represents a concomitant measurement of EELVWI-WO and EELVCT. Horizontal continuous line and horizontal broken lines are the mean bias and 95% prediction interval limits of the bias between EELVWI-WO and EELVCT, respectively. EELVWI-WO, end-expiratory lung volume assessed with the nitrogen washin-washout technique; EELVCT, end-expiratory lung volume assessed by computed tomography; 95% p.i., 95% prediction interval of the bias between EELVWI-WO and EELVCT.

EELV WI - WO - EELV CT = - 1.28 0.33 × EELV WI - WO + EELV CT / 2 .
(1)

Limits of agreement of the bias were ±398 mL (Figure 3), and percentage error was computed to 57%. A significant interaction between PEEP, VT, EELV baseline value on the bias between methods was identified (Table 1) and reported (Figure 4). The bias between methods was strongly influenced by PEEP level, increasing at higher PEEP regardless the VT level. The bias further increased when high PEEP was combined to low VT, when EELV at baseline was low.

The bias at PEEP 0 amounted to -54 ± 101 mL, was not significantly different from 0, and was compared to bias values at higher PEEP. As shown in Figure 5, the bias at PEEP 10 and higher did significantly differ from PEEP 0, while non-significant differences were found for lower PEEP.

Table 1 Statistical modeling of the bias between EELV WI-WO and EELV CT as a function of confounding variables
Figure 4
figure 4

Interaction plot between V T , PEEP, and EELV Base . VT, PEEP, and EELVBase were classified as high or low, as a function of their relationship with their median value. The following cut-off values were identified: VT 170 mL; PEEP 10 cm H2O; EELVBase = 157 mL. Bars are mean values, and error bars, standard deviations. Bias, mean difference between EELVWI-WO and EELVCT in each subgroup; EELVBase, end-expiratory lung volume at baseline (ARDS onset, PEEP 0); EELVWI-WO, end-expiratory lung volume assessed with the nitrogen washin-washout technique; EELVCT, end-expiratory lung volume assessed by computed tomography; PEEP, positive end-expiratory pressure; VT, tidal volume.

Figure 5
figure 5

Mean difference in bias between each PEEP level from 2 to 20 and PEEP 0 cm H 2 O. Closed circles are mean differences and bars are 95% confidence intervals.

Assessment of trending ability of the WI-WO technique

ΔEELVWI-WO values adequately tracked ΔEELVCT changes against time (Figure 6). Concordance rate over all measurements amounted to 79%, and slightly increased after exclusion of small changes in EELV which do not reflect trending ability (Table 2).

Figure 6
figure 6

Four quadrants plot relating ΔEELV WI-WO with ΔEELV CT between consecutive measurements. Continuous black lines are quadrant limits. Dotted line is the regression line. Dashed line is the line of identity. Each data point is the change in end-expiratory lung volume (EELV) between consecutive measurements assessed with the nitrogen washin-washout technique (ΔEELVWI-WO) or computed tomography (ΔEELVCT).

Table 2 Concordance rate, angular bias, and radial limits of agreement in different data subsets

Results of the polar plot analysis are reported in Figure 7, and in Table 2. The angular bias amounted to -4° ± 37° for all measurements. After exclusion of EELV changes ≤100 mL, the angular bias amounted to 3° ± 25°, and was not statistically different from 0. Radial limits of agreement were wide when all measurements were taken into account (±78°), but where narrowed to ±51° after exclusion of EELV changes ≤100 mL. Increasing the exclusion threshold of EELV changes up to 300 mL did not improve the radial limits of agreement (Table 2). Diagnostic performance of EELVWI-WO to detect absolute EELV changes greater than 100, 150, 200, 250, and 300 mL is presented in Table 3. Diagnosis accuracy was fair for detection of absolute EELV changes above 200 mL (AUC 0.79 (CI 95% 0.70 to 0.89)), and good for detection of absolute EELV changes above 300 mL (AUC 0.89 (CI 95% 0.83 to 0.95)).

Figure 7
figure 7

Polar plots assessing trending ability of EELV WI-WO to track changes in EELV. Panel A refers to the whole set of measurements, and panel B is restricted to data related to changes in EELV greater than 100 mL since a small change in EELV does not reflect trending ability but mainly random error measurement. The radial axis joining 0 to 180° is a 45° clockwise rotation of the line of identity in the four-quadrant plots, and represents agreement. The better the agreement between ΔEELV measurements, the closer data pairs will lie along the horizontal radial axis. The distance from the center of each plot represents the mean change in EELV between methods (mean ΔEELV) at each consecutive time point. Data points located between 315° and 45° refer to time points in which both EELVCT and EELVWI-WO increased (upper right quadrant of the four quadrant plot), while data points located between 135° and 225° refer to consecutive time points in which both EELVCT and EELVWI-WO decreased (lower left quadrant of the four quadrant plot). Data points located between 45 and 135° or 225 and 315° correspond to disagreement in the directional change of EELV between the washin-washout technique and computed tomography. Continuous line represents the angular bias, while dashed lines represent radial limits of agreement. EELVWI-WO, end-expiratory lung volume assessed with the nitrogen washin-washout technique; EELVCT, end-expiratory lung volume assessed by computed tomography; ΔEELV, change in EELV between consecutive measurements.

Table 3 Diagnostic performance of EELV WI-WO to detect variations in ELLV CT at different thresholds

Discussion

The main findings of the present study are that, in experimental ARDS, (1) EELVWI-WO underestimates EELVCT, and this underestimation increases linearly as EELV increases; (2) this underestimation is dependent on ventilatory settings (mainly PEEP); (3) the precision of this technique is poor with a percentage error as high as 57%; (4) this technique is however reliable to detect an EELV change greater than 200 mL.

A formal comparison between EELVWI-WO EELVCT has already been performed in a pig model of pleural effusion [19] and in mechanically ventilated patients [2]. While Chiumello et al. found a slightly positive constant bias between EELVWI-WO and EELVCT[2], and Graf et al. a slightly negative constant bias, a non-constant linear bias was found in the present study. Beside differences in species, experimental protocol or mechanism of lung injury, the likely explanation of this discrepancy is related to the higher PEEP level applied in the present study (up to 20 cm H2O), while PEEP was set to 5 cm H2O in the human study [2] and ≤10 cm H2O in the animal study [19]. The dependence of bias on PEEP level was further emphasized in our study by multivariate analysis (Table 1, Figure 4), more specifically for PEEP levels greater than 10 cm H2O. Finally, a detrimental effect of PEEP levels greater than 16 cm H2O on EELVWI-WO accuracy was also recently pointed out by Dellamonica et al. [23]. Nevertheless, in spite of a non-constant bias, the true EELV value may still be assessed by the WI-WO technique using Equation 1, as pointed out by Bland and Altman [24].

The detrimental effect of high PEEP levels on the bias may have multiple explanations. First, the WI-WO technique measures volume of lung regions that may be reached by nitrogen (and hence are ventilated), while CT measures aeration of both ventilated and non-ventilated regions. One could then hypothesize that regional ventilation (and hence nitrogen during EELV measurement) at high PEEP may be preferentially directed toward non-overinflated regions and that overinflated regions may not be detected in the EELV measurement. However, this explanation is not supported by our data, since the amount of overinflated area was very low even at the highest PEEP level in our study (1.1 ± 0.89 mL). Nevertheless, the observed difference in EELV between methods at high PEEP may be related to the fact that the WI-WO method measures a functional EELV, while CT measures anatomical EELV. Occurrence of leaks at high PEEP is another hypothesis to explain the non-constant bias, but this explanation is unlikely since plateau pressure was maintained during the prolonged end-inspiratory pauses performed for CT acquisition. Finally, there may be insufficient nitrogen mixing within the aerated lung, during the time allocated for measurement by the ventilator, since animals were ventilated at high RR and low VT, which may have increased the time to reach equilibrium during the WI-WO measurement. An extended time between experimental stages and/or a lengthened WI-WO period may have narrowed the bias between methods.

Regarding precision, comparison between aforementioned studies, using limits of agreement is hindered by heterogeneity of EELV values across studies. Assessment of percentage error may overcome this problem, but was unavailable in the two previously published studies [2, 19]. Using Data Thief 3.0 to extract Cartesian data from these studies, we have computed from the author’s published figures, the percentage error which amounted to 28% in the human study and 46% in the pig study, versus 57% in our study. The relatively lower precision computed from our data may be a consequence of ventilatory settings, that may have particularly challenged the validity of the WI-WO technique. Indeed, FiO2 greater than 0.7 precludes the computation of the respiratory quotient (RQ) required for EELV measurements, and a default RQ of 0.85 is assumed by the Engström Carestation® ventilator.

However, using a metabolically active lung model, Olegard et al. have nicely demonstrated that errors in RQ computation have a negligible effect on the precision of EELV measurements [1], suggesting that the high FiO2 used in our study may have only marginally influenced our results. Regarding trending, the lack of significant angular bias suggests that calibration of EELVWI-WO is in agreement with the reference method. However, the relatively wide radial limits of agreement suggest that external factors may account for the variability of the relationship between ΔEELVWI-WO and ΔEELVCT. We could speculate that this phenomenon is mainly related to the effect of PEEP on the bias between EELVWI-WO and ΔEELVCT.

Our study has several strengths. Since multiple combinations of PEEP and VT were evaluated, a systematic analysis of the effect of ventilator parameters on the reliability of the technique could be performed using multivariate analysis, and was able to identify the PEEP level as an independent risk factor for measurement error. Furthermore, this is the first study having assessed trending ability of the WI-WO technique, and provided with cut-off values above which EELV changes may be considered as meaningful. Finally, the present study, while performed with particularly challenging ventilatory settings (high FiO2, high PEEP, high RR, and low VT), demonstrates that the validity of the WI-WO technique may be extended to the sickest ARDS patients.

Our study has nevertheless some methodological issues that must be addressed. First, it was conducted using pediatric sensors for EELVWI-WO measurements, which may be less accurate for the highest VT. Nevertheless, a subset analysis after exclusion of VT > 300 mL led to similar results (Additional file 3: Figure S1). Furthermore, the bias was not increased at high VT, as compared to low VT, as shown by our interaction plot (Figure 4). Another limitation, in the perspective of extrapolation of these results to ARDS patients, is related to the relatively low EELV achieved in the pigs of this study in some experimental conditions. However, 62% of the EELVCT measurements were greater than the first EELV quartile observed at low PEEP in a recent study on 30 ARDS patients [23], suggesting that most of the measurements performed in our study are in the range of clinically plausible values for EELV in ARDS patients. Another potential limitation is that the ARDS model used in the present study is particularly recruitable with PEEP. However, 50% of ARDS patients are considered recruiters by PEEP [23, 25], and early ARDS share similar features as saline lavage regarding response to PEEP. The 2-min interval between VT changes and EELV measurements may have been too short for CO2 equilibration and achievement of both progressive recruitment and blood flow redistribution. However, a subset analysis limited to data acquired during the PEEP trial (with 10 min between measurements) led to similar results (see Additional file 4: Figure S2).

The present study may have important clinical implication. Indeed, as shown in Table 3, a change in EELVWI-WO greater than 166 mL would give an important clue to the clinician that the true EELV has changed by more than 200 mL. Furthermore, despite the non-constant bias of the EELV measurement by the WI-WO technique, the true EELV value may still be assessed using Equation 1, provided that the absolute EELV value is relevant for the clinician [24].

Conclusion

The reliability of the WI-WO technique is critically dependent on ventilatory settings, but sufficient to accurately detect EELV change over time greater than 200 mL.

Abbreviations

ARDS:

acute respiratory distress syndrome

AUC:

area under curve

ΔEELVWI-WO:

change in end-expiratory lung volume between consecutive measurements by the nitrogen washin-washout technique

ΔEELVCT:

change in end-expiratory lung volume between consecutive measurements by computed tomography

EELV:

end-expiratory lung volume

EELVWI-WO:

end-expiratory lung volume measurement by the nitrogen washin-washout technique

EELVCT:

end-expiratory lung volume measurement by computed tomography

FiO2:

fraction of inspired oxygen

PEEP:

positive end-expiratory pressure

RR:

respiratory rate

SD:

standard deviation

V T :

tidal volume

WI-WO:

nitrogen washin-washout.

References

  1. Olegard C, Sondergaard S, Houltz E, Lundin S, Stenqvist O: Estimation of functional residual capacity at the bedside using standard monitoring equipment: a modified nitrogen washout/washin technique requiring a small change of the inspired oxygen fraction. Anesth Analg 2005, 101: 206–212. 10.1213/01.ANE.0000165823.90368.55

    Article  PubMed  Google Scholar 

  2. Chiumello D, Cressoni M, Chierichetti M, Tallarini F, Botticelli M, Berto V, Mietto C, Gattinoni L: Nitrogen washout/washin, helium dilution and computed tomography in the assessment of end expiratory lung volume. Crit Care 2008, 12: R150. 10.1186/cc7139

    Article  PubMed  PubMed Central  Google Scholar 

  3. Dellinger RP, Levy MM, Rhodes A, Annane D, Gerlach H, Opal SM, Sevransky JE, Sprung CL, Douglas IS, Jaeschke R, Osborn TM, Nunnally ME, Townsend SR, Reinhart K, Kleinpell RM, Angus DC, Deutschman CS, Machado FR, Rubenfeld GD, Webb S, Beale RJ, Vincent JL, Moreno R, Surviving Sepsis Campaign Guidelines Committee including The Pediatric Subgroup: Surviving sepsis campaign: international guidelines for management of severe sepsis and septic shock, 2012. Intensive Care Med 2013, 39: 165–228. 10.1007/s00134-012-2769-8

    Article  PubMed  CAS  Google Scholar 

  4. Sondergaard S, Karason S, Wiklund J, Lundin S, Stenqvist O: Alveolar pressure monitoring: an evaluation in a lung model and in patients with acute lung injury. Intensive Care Med 2003, 29: 955–962.

    PubMed  CAS  Google Scholar 

  5. Top A, Hamarneh G, Abugharbieh R: Spotlight: Automated Confidence-Based user Guidance for Increasing Efficiency in Interactive 3D Image Segmentation. In Medical Computer Vision. Recognition Techniques and Applications in Medical Imaging. Edited by: Menze B, Langs G, Tu Z, Criminisi A. Berlin Heidelberg: Springer Berlin Heidelberg; 2011:204.

    Chapter  Google Scholar 

  6. Top A, Hamarneh G, Abugharbieh R: Active Learning for Interactive 3D Image Segmentation. In Medical Image Computing and Computer-Assisted Intervention – MICCAI 2011. Edited by: Fichtinger G, Martel A, Peters T. Heidelberg: Springer Berlin; 2011.

    Google Scholar 

  7. Attinger EO, Cahill JM: Cardiopulmonary mechanics in anesthetized pigs and dogs. Am J Physiol 1960, 198: 346–348.

    PubMed  CAS  Google Scholar 

  8. Acute Respiratory Distress Syndrome Network: Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N Engl J Med 2000, 342: 1301–1308.

    Article  Google Scholar 

  9. R Core Team: R: a Language and Environment for Statistical Computing. Vienna, Austria: R Foundation for statistical Computing; 2013. http://cran.r-project.org

    Google Scholar 

  10. Pinheiro J, Bates D, Debroy S, Sarkar D, The R Development Core Team: Nlme: Linear and Nonlinear Mixed Effects Models. 2013. http://cran.r-project.org/web/packages/nlme/index.html

    Google Scholar 

  11. Carstensen B, Gurrin L, Ekstrom C, Figurski M: MethComp: Functions for Analysis of Agreement in Method Comparison Studies. 2013. http://cran.r-project.org/web/packages/MethComp/index.html

    Google Scholar 

  12. Robin X, Turck N, Hainard A, Tiberti N, Lisacek F, Sanchez J-C, Müller M: pROC: an open-source package for R and S + to analyze and compare ROC curves. BMC Bioinforma 2011, 12: 77. 10.1186/1471-2105-12-77

    Article  Google Scholar 

  13. Lopez-Raton M, Rodriguez-Alvarez MX: Optimal Cutpoints: Computing Optimal Cutpoints in Diagnostic Tests. 2012. http://cran.r-project.org/web/packages/OptimalCutpoints/index.html

    Google Scholar 

  14. Hothorn T, Bretz F, Westfall P: Simultaneous inference in general parametric models. Biom J 2008, 50: 346–363. 10.1002/bimj.200810425

    Article  PubMed  Google Scholar 

  15. Bland JM, Altman DG: Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat 2007, 17: 571–582. 10.1080/10543400701329422

    Article  PubMed  Google Scholar 

  16. Carstensen B: Comparing Clinical Measurement Methods: a Practical Guide (Statistics in Practice). 2nd edition. Chichester: Wiley; 2011.

    Google Scholar 

  17. Carstensen B, Simpson J, Gurrin LC: Statistical models for assessing agreement in method comparison studies with replicate measurements. Int J Biostat 2008, 4: 1–26.

    Article  Google Scholar 

  18. Critchley LA, Critchley JA: A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 1999, 15: 85–91. 10.1023/A:1009982611386

    Article  PubMed  CAS  Google Scholar 

  19. Graf J, Santos A, Dries D, Adams AB, Marini JJ: Agreement between functional residual capacity estimated via automated gas dilution versus via computed tomography in a pleural effusion model. Respir Care 2010, 55: 1464–1468.

    PubMed  Google Scholar 

  20. Tummers B: Datathief III. 2006. http://datathief.org

    Google Scholar 

  21. Critchley LA, Lee A, Ho AM-H: A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg 2010, 111: 1180–1192. 10.1213/ANE.0b013e3181f08a5b

    Article  PubMed  Google Scholar 

  22. Critchley LA, Yang XX, Lee A: Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth 2011, 25: 536–546. 10.1053/j.jvca.2011.01.003

    Article  PubMed  Google Scholar 

  23. Dellamonica J, Lerolle N, Sargentini C, Beduneau G, Di Marco F, Mercat A, Richard JC, Diehl JL, Mancebo J, Rouby JJ, Lu Q, Bernardin G, Brochard L: PEEP-induced changes in lung volume in acute respiratory distress syndrome. Two methods to estimate alveolar recruitment. Intensive Care Med 2011, 37: 1595–1604. 10.1007/s00134-011-2333-y

    Article  PubMed  CAS  Google Scholar 

  24. Bland JM, Altman DG: Measuring agreement in method comparison studies. Stat Methods Med Res 1999, 8: 135–160. 10.1191/096228099673819272

    Article  PubMed  CAS  Google Scholar 

  25. Gattinoni L, Caironi P, Cressoni M, Chiumello D, Ranieri VM, Quintel M, Russo S, Patroniti N, Cornejo R, Bugedo G: Lung recruitment in patients with the acute respiratory distress syndrome. N Engl J Med 2006, 354: 1775–1786. 10.1056/NEJMoa052052

    Article  PubMed  CAS  Google Scholar 

Download references

Role of the sponsors

The study was founded by General Electric who provided with the ventilator and disposables, technical assistance, and funded the study (9,600 €). The investigators had no ties with this company. The study was also funded by grant no. C11S01 from the French-Colombian program ECOS-Nord. These sponsors had no role in the design, collection, analysis, and interpretation of data; in the writing of the manuscript; and in the decision to submit the manuscript for publication. AMP was supported by a Colombian doctoral grant from Colciencias and by a Rhône-Alpes Region grant CMIRA.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Claude Guerin.

Additional information

Competing interests

JCR, CP, AMP, JST, MO, BN, MHH, and FL have no competing interests. CG received a grant of 9600 € by GE Healthcare as study funding. CG has no other ties with this company, and the manuscript was submitted without review by this company.

Authors’ contributions

JCR and CG have made substantial contributions to the conception, design, acquisition of data, and analysis and interpretation of data; AMP, JST, MO, BN, and MHH to the analysis and interpretation of data; MO to the conception, design, and analysis and interpretation of data; CP to the conception, acquisition of data, and analysis and interpretation of data; and FL to the acquisition of data. JCR has drafted the submitted article. CP, AMP, JST, MO, BN, MHH, FL, and CG have revised the draft critically for important intellectual content. All authors have provided the final approval of the version to be published; and agreed to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.

Electronic supplementary material

40635_2013_12_MOESM1_ESM.docx

Additional file 1: Table S1: Reasons for lack of data as a function of each of the 3 experimental stages. Values are number of lacking data/total number of data (%). EELVWI-WO, end-expiratory lung volume assessed with the nitrogen washout-washin technique; EELVCT, end-expiratory lung volume assessed by computed tomography; VT, tidal volume. (DOCX 15 KB)

40635_2013_12_MOESM2_ESM.docx

Additional file 2: Table S2: Ventilatory settings and arterial blood gases in each experimental condition. Values are number of mean ± standard deviation (range). ALI, acute lung injury onset; PEEP, positive end-expiratory pressure; RR, respiratory rate; VT, tidal volume. (DOCX 17 KB)

40635_2013_12_MOESM3_ESM.docx

Additional file 3: Figure S1: Bias and limits of agreement between EELVCT and EELVWI-WO, using Bland and Altman representation in a subset of the data (exclusion of VT > 300 mL). Each symbol represents a concomitant measurement of EELVWI-WO and EELVCT. Horizontal continuous line and horizontal broken lines are the mean bias and 95% prediction interval limits of the bias between EELVWI-WO and EELVCT, respectively. EELVWI-WO, end-expiratory lung volume assessed with the nitrogen washout-washin technique; EELVCT, end-expiratory lung volume assessed by computed tomography; 95% p.i., 95% prediction interval of the bias between EELVWI-WO and EELVCT; VT, tidal volume. (DOCX 278 KB)

40635_2013_12_MOESM4_ESM.docx

Additional file 4: Figure S2: Bias and limits of agreement between EELVCT and EELVWI-WO, using Bland and Altman representation in a subset of the data, acquired during the PEEP trial (with 10 min between measurements). Each symbol represents a concomitant measurement of EELVWI-WO and EELVCT. Horizontal continuous line and horizontal broken lines are the mean bias and 95% prediction interval limits of the bias between EELVWI-WO and EELVCT, respectively. EELVWI-WO, end-expiratory lung volume assessed with the nitrogen washin-washout technique; EELVCT, end-expiratory lung volume assessed by computed tomography; 95% p.i., 95% prediction interval of the bias between EELVWI-WO and EELVCT; VT, tidal volume. (DOCX 137 KB)

Authors’ original submitted files for images

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 2.0 International License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

Reprints and permissions

About this article

Cite this article

Richard, JC., Pouzot, C., Pinzón, A.M. et al. Reliability of the nitrogen washin-washout technique to assess end-expiratory lung volume at variable PEEP and tidal volumes. ICMx 2, 10 (2014). https://doi.org/10.1186/2197-425X-2-10

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/2197-425X-2-10

Keywords